Calibration procedure, Single detector – INFICON Explorer Portable Gas Chromatograph User Manual
Page 223
Explorer Operator Manual
221
Calibration Procedure
Single Detector
1. Select a column.
2. Press Enter Menu, then SETUP, CONFIG, and COLUMN.
3. Use the arrow keys to move through the list and press Enter Menu to select
PID Column A, B or C or ECD Column A, B or C.
Ensure only one column/detector combination is selected.
There are two libraries for each column. The PID library or the ECD library is
automatically selected when you choose a column/detector combination.
4. Press Exit to close the box.
5. Press Enter Menu, then LIBRARY, and LIST.
6. Press the left/right arrow to select the library that matches your selected
column.
The status line at the top of the screen will tell you which library you are using.
Compounds with a check mark are currently selected for calibration.
7. Use the up/down arrows to move through the list of compounds, and then to
enable or disable a compound, move to the compound name and press EDIT
and CALCMPD.
This will toggle the compound on and off. If the compound is not selected
remove it from the library. If the compound is not selected, its response and
retention time will be updated ratio-metrically.