Library – INFICON Explorer Portable Gas Chromatograph User Manual
Page 214

Calibration
212
Library
After each analysis,
Explorer compares detected peaks against those in the library.
It identifies those that match library compounds and then quantifies the identified 
peaks using the sensitivity from the library. Peaks are identified by comparing the 
retention time of the detected peak to the retention times stored in the library, or it 
identifies the peak as the largest peak within a retention time window. 
Explorer calculates a sensitivity value for each compound when it is stored in the
library. The sensitivity is the ratio of the measured response (peak area) to the actual 
concentration of the compound. Each time a calibration is performed the retention 
time and sensitivity values of each compound are updated in the library. 
Each assay includes a library of compounds. The compounds and all associated 
calibration information have been provided as part of the assay. You may run the 
assay as it has been programmed for you, or you may customize the assay for your 
particular application. 
If you have more than one assay or if you have created your own assay, a calibration 
must be performed to update the library information each time a different assay 
library is loaded for use. 
If
Explorer will be set up for loop injection, compounds must be stored by loop
injection and then calibrations must be performed using loop injection 
techniques. Samples that are injected manually by syringe must be stored by 
syringe and then calibration must be performed by syringe injection. If both loop 
and syringe injection techniques are used, create a library for each. 
NOTE: Do not use a library created by loop injection interchangeably with one created by
syringe injection.
