INFICON Explorer Portable Gas Chromatograph User Manual
Page 156
SiteChart
Software
154
Remote Connection Hardware Configuration
This option allows you to setup
Explorer at the monitoring site. You will setup the
computer at a remote location and communicate with the
Explorer through a
modem. You must have a modem for
Explorer and another modem for the
computer. A modem capable of transmitting at 9600 baud is required. A faster
modem is recommended.
To configure your system for remote communication:
1. Connect the modem to the
Explorer using the communication cable (Part No.
MX380304
with 4253000).
2. Turn
Explorer on and ensure carrier gas is connected or that the internal
reservoir is full.
3. Connect the modem to the phone line.
4. Ensure the modem is turned on and is working properly.
5. Connect the second modem to a serial port on the computer.
If you need to add another serial port, refer to the computer’s manual for
detailed instructions.
NOTE: The modem must be connected to a serial port.
6. In the Communication menu, select
Setup
.
The Setup command allows you to configure the computer to communicate
with
Explorer either directly or through a modem.
7. Select the serial port to which the modem is connected.
8. Connect the modem to the phone line.
9. Ensure the modem is turned on and is working properly.