Updating Library Information
If your chromatographic settings have not changed from the original assay, but
library compounds are not correctly identified or not identified at all, they may have
drifted out of the peak recognition window. You can update library information
based on a recent chromatogram that has been run using the original assay.
To update library information:
1. Select the Assay tab.
2. Open the original assay and save it with a new name.
3. Click
Send to
Explorer
to send the assay information to
Explorer.
4. Select the first column/detector pair.
If your application uses 2 or 3 columns you must repeat this procedure for each
column. If you are using the ECD, you must repeat this procedure for each
ECD library.
5. Remove all the unwanted compounds from the library.
To remove a compound from the library, select it and press
Delete
.
6. Run a chromatogram on the selected column/detector pair.
The chromatogram must contain all the compounds listed in your library. It
must also be run using your original assay method.
Allow
Explorer to warm up for at least 30 minutes before running the
chromatogram.
7. Re-integrate the chromatogram.
Use the
re-integrate
command in the Analyzer menu or right-click on the
chromatogram and select
re-integrate
from the pop-up menu. Compounds
matching library information will be identified.
8. Click on the first compound in the library.