5.11. Setting Up Digital Inputs

The Protocol 3 controller can be run by external inputs wired to the controller from an external
source such as a PLC or switches on a control panel (Figure 21 displays standard digital input
wiring).



















Enable Digital Inputs through the

Profile Menu

, then

General Configuration

and set

Remote

Profile Control

to

Enable

.


While the Remote Profile Control is enabled, profiles can be run
only by digital inputs (versus running profiles from the Profile
Mode).



The external run operation can

Run

,

Hold

or

Stop

profiles indexed in positions 1-7 in the

memory of the control. Table 19 shows the profile selection.

Table 19. Input settings.
Input 1

Input 2

Input 3

Index Selection

ON

OFF

OFF

1

OFF

ON

OFF

2

ON

ON

OFF

3

OFF

OFF

ON

4

Figure 21. Standard Digital Input Wiring.