Alarm Selection
Parameter 1
Parameter 2
Parameter 3
Process Low
Alarm Value
(Degrees)
Alarm Hysteresis (Degrees)
Alarm Inhibit
PV-SP Deviation
Alarm Value
(Degrees)
Alarm Hysteresis (Degrees)
Alarm Inhibit
Band
Alarm Value
(Degrees)
Alarm Hysteresis (Degrees)
Alarm Inhibit
Signal Change Per
Min
Alarm Value
(Degrees)
Minimum Duration
(Seconds)
Alarm Inhibit
Input Signal Break
N/A
N/A
N/A
Control Loop
Loop Alarm
Type
Manual Loop Alarm Time
(mm:ss) (active only if loop
alarm type is manual)
N/A
% Memory Used
Alarm Value (%) Alarm Inhibit
N/A
5.6.1.3. Using Alarms to Switch Relay Outputs
The Protocol 3 controller allows alarm states to trigger any of the five relay outputs on the
controller. Each output can be set to a specific alarm or multiple alarms using the
Or Alarms
and
Alarm and Events
functions. To configure an alarm state to trigger a relay output:
1. From
Configuration Menu
, use
and
to scroll to and highlight
Output
Configuration
then press
to access the
Output Configuration
Menu
.
2. Press
to move forward through the different parameters in the menu,
and
to enter the setting for a parameter,
to move back through the different parameters.
3. Press
and
together (simultaneously) to return to the
Configuration Menu
.