5.6.2. Working with Calibration Offsets

Calibration offsets allow configuration for oven temperature variations. Enter calibration offsets
through the

Configuration Menu

and

Input Configuration

.

5.6.2.1. Access the Configuration Menu

To access the

Configuration Menu

and all sub-menus, the controller must be in its Cycle

Complete or Stopped state (that is, not running). If the controller is not in the Stopped state, press

to stop the controller before proceeding.

1. Press

and

together from the

Select a Mode

screen to access the Main Menu.

2. From the Main Menu, navigate to and select (highlight) the

Configuration Menu

. Press

.

3.

A prompt will appear to enter an unlock code. Enter the correct code (the default code is

0010) with

and

then press

to access the

Configuration Menu

.



5.6.2.2. Working with Input Configuration

Input Configuration

allows you to configure items associated with the

Control Input Setup

,

Control Input Cal.

,

High Limit Input Setup

and

High Limit Input Cal

.

Control calibration offset

parameters are located in the

Control Input Cal

.

To work with

Input Configuration

:

1. Navigate from the

Configuration Menu

to and highlight

Input Configuration

. Press

.

2. Navigate to and select

Control Input Cal.

Press

.

a.

Three options present:

Factory Calibration

,

Single Point Calibration

,

Two Point

Calibration

.

b. Only one option can be active at any given time. The highlighted choice becomes the

active choice.

3. Navigate to and select the desired choice. Press

to configure the settings within

that choice.