6 thermocouple replacement -11, 6 thermocouple replacement – Bacharach PCA 2 24-9448 User Manual
Page 91

Instruction 24-9448
6-11
Maintenance
6.6 Thermocouple Replacement
Usng the approprate thermocouple replacement kt lsted below, replace
the probe’s thermocouple as follows:
Thermocouple Replacement Kits:
Part Number Replaces Thermocouple in a
Probe with a Tube Length of . . .
24-8413
6 nches
24-8414
12 nches
24-8415
24 nches
24-8416
36 nches
Each kt contans a thermocouple assembly, two O-rngs, and two wre-
splce connectors.
Tools Required:
• Small Flat Blade Screwdrver
• Wre Cutter
• Wre Strpper
• Slp Jont Plers
Procedure:
1. Gan access to the thermocouple connectons by frst removng three
screws from probe handle, and then separatng the two handle peces.
2. Cut wres attached to old crmp connectors, leavng behnd as much of
the probe’s thermocouple-connector wre as possble.
Probe Handle
Screws
Figure 6-6. Probe Handle Disassembly