beautypg.com

2 turnng on the analyzer and warm up -2, 2 turning on the analyzer and warm up – Bacharach PCA 2 24-9448 User Manual

Page 36

background image

Instruction 24-9448

4-2

Operation

4.2 Turning ON the Analyzer and Warm Up

1. Connect the probe and hose assembly, and make sure that the analyzer

s properly set up per Secton 3 Initial Setup.

IMPORTANT: DO NOT insert probe into stack

before turning ON the analyzer!

2. Place the probe n an area that contans fresh

ar. Ths ensures that the sensors wll be

properly zeroed durng the warm-up cycle.

3. Turn ON the analyzer by pressng the I/O

button for at least 1 second, or untl a sngle

beep s heard. Observe that the analyzer’s

frmware verson, model and seral numbers

are brefly dsplayed followed by the Warm Up

screen.

4. Wat for the analyzer to count down ts

60 second warm-up perod; after whch, the

message “NO ERRORS DETECTED” along

wth the oxygen sensor’s output voltage s

brefly dsplayed followed by the dsplay of the

Combuston Test HOLD screen.

TIP: The O

2

value at the end of warm-up is

reported as "Good, Low or BAD". Consider

replacing the oxygen sensor when its value is

reported as "Low".

If problems were detected durng warm up,

the message “ERRORS DETECTED” s

dsplayed along wth a lst of those errors.

As an example, the screen to the rght shows

that the battery s low. Refer to Secton 7.3 for

a lstng and possble remedy for the errors

dsplayed.

TIP: If the sensors in error are not critical to

the combustion test, then press the RUN/HOLD

button to display the Combustion Test HOLD

screen and proceed with the test.

MENU

Warm Up 60

MENU

Warm Up Complete

NO ERRORS DETECTED

MENU

O

2

Value: Good

ERRORS DETECTED
Low Battery

MENU

F1

F2

F3

MENU

MENU

PCA 2

FIRMWARE VERSION

x.xx

MODEL xxx

SERIAL # xxxxxxx