beautypg.com

Agilent Technologies Switch/Measure User Manual

Page 71

background image

Features and Functions

2

34980A User’s Guide

59

Front Panel Operation:

View > READINGS

Remote Interface Operation:

The following command retrieves stored

readings from memory (the readings are not erased).

FETCh?

Use the following commands to query the statistics on the readings stored
in memory for a specific channel or from the internal DMM. These
commands do not remove the data from memory.

The following command retrieves the last reading taken on channel 1 on
the module in slot 3 during a scan.

DATA:LAST? (@3001)

The following command clears the contents of statistics memory for the
selected channel.

CALC:AVER:CLEAR (@3001)

Use the following command to determine the total number of readings
stored in memory (all channels) from the most recent scan.

DATA:POINTS?

The following command reads and clears the specified number of readings
from memory. This allows you to continue a scan without losing data
stored in memory (if memory becomes full, new readings will overwrite
the first readings stored). The specified number of readings are cleared
from memory, starting with the oldest reading.

DATA:REMOVE? 12

CALC:AVER:MIN? (@3005)

CALC:AVER:MIN:TIME? (@3005)

CALC:AVER:MAX? (@3005)

CALC:AVER:MAX:TIME? (@3005)

CALC:AVER:AVER? (@3005)

CALC:AVER:COUNT? (@3005)

CALC:AVER:PTPEAK? (@3005)

Minimum reading on channel
Time minimum was logged

Maximum reading on channel
Time maximum was logged

Average of all readings on channel

Number of readings taken on channel

Peak- to- peak (maximum–minimum)