34932a dual 4x16 armature matrix – Agilent Technologies Switch/Measure User Manual
Page 161
![background image](/manuals/41536/161/background.png)
Matrix Switch Modules
5
34980A User’s Guide
149
34932A Dual 4x16 Armature Matrix
The 34932A dual 4x16 armature matrix contains two matrices, each with 64
2- wire crosspoint latching armature relays organized in a 4- row by 16- column
configuration. Every row and column are made up of two wires each, a high
(H) and a low (L). Each crosspoint relay has a unique channel number
representing the row and column that intersect to create the crosspoint. For
example, channel 315 represents the crosspoint connection between row 3
and column 15 (all columns consisting of two digits; in this case the digits are
15). See the simplified schematic on
.
You can connect any combination of inputs and outputs at the same time.
However, only Matrix 2 in this module connects to the Analog Buses. By
closing channels 921 and 922 you can connect rows 5 and 6 respectively to
the internal DMM of the 34980A mainframe for voltage and resistance
measurements. You can connect multiple matrix modules externally and/or
through the Analog Buses for applications that require large matrices. For
information on linking multiple matrix modules, refer to
of this
chapter.
When the power is off, matrix relays maintain state, and Analog Bus
relays open.
N O T E
When the DMM is scanning, it controls ABus1 and ABus2 relays,
which are on Matrix 2. Therefore, consider this behavior when you
are connecting matrices.