34980A User’s Guide
2
Features and Functions
Front Panel Operation:
• To configure the measurement parameters and add a channel to the
scan list, use the
Channel (Configure) key.
• To initiate a scan and store all readings in memory, press the
Scan (Measure) key. If you press the Scan (Measure) key with no
scan list defined, the instrument initiates a DMM- only measurement
(see
below).
• To stop a scan in progress, press and hold the
Scan (Measure) key.
• To view the readings in memory, use the
View
key (the readings are not
erased when you read them).
Remote Interface Operation:
• To define the list of channels to be included in the scan list, use the
ROUTe:SCAN
command.
• To configure the measurement parameters on the desired channels,
use the
CONFigure
and
SENSe
commands.
• To initiate a scan and store all readings in memory, use the
INITiate
or
READ?
command. Each time you initiate a new scan, the instrument
will clear the previous set of readings from memory.
• To stop a scan in progress, use the
ABORt
command.
• To view the readings in memory, use the
FETCh?
command
(the readings are not erased when you read them).
N O T E
You can use the READ? command in one of three forms depending on
which measurement mode you wish to use.
•
If you omit the optional
currently defined, the READ? command applies to the internal DMM.
•
If you omit the optional
defined, the READ? command performs a scan of the channels in the
scan list.
•
If you specify a
defined, the READ? command performs a “temporary” scan of the
specified channels (independent of the present scan list).
N O T E
You can use the MEASure? command in one of two forms depending on
which measurement mode you wish to use.
•
If you omit the optional
applies to the internal DMM.
•
If you specify a
“temporary” scan of the specified channels (independent of the
present scan list).