Agilent Technologies Switch/Measure User Manual

Page 32

background image

20

34980A User’s Guide

2

Features and Functions

The following table shows the relationship between integration time,

measurement resolution, number of digits, and number of bits.

The instrument selects 1 PLC when the measurement function is

changed and after a Factory Reset (

*RST

command). An Instrument

Preset (

SYSTem:PRESet

command) or Card Reset (

SYSTem:CPON

command) does not change the integration time setting.

Front Panel Operation:

DMM or Channel (Configure) > INTEGRATION > TIME

First, select the measurement function on the active channel. You are
automatically guided to the next level of the menu where you can select
a specific integration time.

Remote Interface Operation:

You can set the integration time using the

SENSe

commands. For example, the following command specifies an

aperture time of 2 ms for resistance measurements on channel 2001.

SENS:RES:APER 0.002,(@2001)

Relationship between integration time, resolution, digits, and bits

Integration Time

Resolution

Digits

Bits

0.02 PLC
0.2 PLC
1 PLC
2 PLC
10 PLC

20 PLC
100 PLC
200 PLC

< 0.0001 x Range
< 0.00001 x Range
< 0.000003 x Range
< 0.0000022 x Range
< 0.000001 x Range

< 0.0000008 x Range
< 0.0000003 x Range
< 0.00000022 x Range

4½ Digits
5½ Digits
5½ Digits
6½ Digits
6½ Digits

6½ Digits
6½ Digits
6½ Digits

15
18
20
21
24

25
26
26