Features and Functions

2

34980A User’s Guide

13

To view the readings in memory, use the

View

key (the readings are

not erased when you read them). Each time you initiate a new
DMM- only scan, the instrument will clear the previous set of readings
from memory.

Remote Interface Operation:

You can use the

MEASure?

command without specifying a to

quickly take a stand- alone DMM reading. Note, however, that with the

MEASure?

command, most measurement parameters are set to their

default values.

To close the desired channel relays and Analog Bus relays, use the

ROUTe:CLOSe

command. The Analog Bus relays on the multiplexer and

matrix modules are numbered s911, s912, s913, etc.

To directly control all measurement parameters or triggering, use the

CONFigure

,

SENSe

, and

TRIGger

commands without specifying a

parameter. To initiate the measurement, use the

INITiate

or

READ?

command without specifying a . Each time you initiate a

new measurement, the instrument will clear the previous set of
readings from memory.

To stop a measurement in progress, use the

ABORt

command.

To view the readings in memory, use the

FETCh?

command

(the readings are not erased when you read them).

Scanning Mode

In the Scanning Mode, the 34980A automatically controls a sequence
of measurements using the internal DMM, possibly across multiple
channels, and stores the results in memory. The 34980A closes and
opens the appropriate channel relays and Analog Bus relays required
for the sequence. The following general rules apply to the Scanning Mode
(for more information on using the Scanning Mode, see

“Scanning”

on

page 42.)

Any channel that can be “read” by the instrument can also be included

in a scan. A scan can also include a read of a digital channel or a read
of the totalizer count on the digital modules.

Before you can initiate a scan, you must set up a scan list to include all

desired multiplexer or digital channels. Channels which are not in the
scan list are skipped during the scan.

The Analog Bus relays are automatically opened and closed as required

during the scan to connect to the internal DMM for the measurement.
For example, all 2- wire measurements use the ABus1 (MEAS) relays; for
4- wire measurements, the ABus2 (SENS) relays are used in addition to
the ABus1 relays.

Each time you initiate a new scan, the instrument will clear the

previous set of readings from memory.