Electrical operating conditions – Agilent Technologies Switch/Measure User Manual

Page 105

background image

Introduction to the Plug-In Modules for the 34980A

3

34980A User’s Guide

93

Electrical Operating Conditions

34937A

28 channels, 300 V rms or DC,
1 A, 60 VA per channel
4 channels, 250 V rms or 30 VDC,
5A, 150 VA per channel

28 channels, 100 V rms or DC,
1 A, 60 VA per channel
4 channels, 100 V rms or 30 VDC,
5A, 150 VA per channel

34938A

20 channels, 250 V rms or 30 VDC,
5 A, 150 VA per channel

20 channels, 100 V rms or 30 VDC,
5 A, 150 VA per channel

34941A

Four channels, 30 V, 0.5 A,

10 W per channel

Four channels, 30 V, 0.5 A,

10 W per channel

34946A

Dual channel, 7 V, 1 W per channel,
4 GHz or 20 GHz

Dual channel, 7 V, 1 W per channel, 4
GHz or 20 GHz

34947A

Triple channel, 7 V, 1 W per channel,
4 GHz or 20 GHz

Triple channel, 7 V, 1 W per channel,
4 GHz or 20 GHz

34951A

4 channels, 16 V, 20 mA

4 channels, 16 V, 20 mA

34952A

32 DIO channels, 42 V, 400 mA,
2 channel DAC, 12 V, 10 mA

32 DIO channels, 42 V, 400 mA,
2 channel DAC, 12 V, 10 mA

N O T E

Pollution Degree 1: No pollution or only dry, non-conductive pollution
occurs. The pollution has no influence (on insulation) (IEC 61010-1
2nd Edition).

N O T E

Pollution Degree 2: Normally only non-conductive pollution occurs.
Occasionally, a temporary conductivity (leakage current between
isolated conductors) caused by condensation can be expected (IEC
61010-1 2nd Edition).

C A U T I O N

For proper module cooling, all unused slots must be covered.

Module

Pollution Degree 1 Specifications

Pollution Degree 2 Specifications

WA RN I N G

To avoid electric shock, turn off the 34980A and disconnect or
de-energize all field wiring to the modules and the Analog Bus
connector before removing any module or slot cover.