beautypg.com

Installing sma connectors – Agilent Technologies Switch/Measure User Manual

Page 195

background image

Dual/Triple Microwave Switch Modules

8

34980A User’s Guide

183

Example: Querying the system for module identify

The following command

returns the identify of the module installed in slot 7.

SYSTem:CTYPe? 7

Example: Reading the cycle count for a relay

The following command reads

back the number of completed cycles for the channel 201 relay of a module
installed in slot 6.

DIAGnostic:RELay:CYCLes? (@6201)

Example: Clearing the cycle count for a relay

The following command resets

the cycle count on the channels 201 and 202 for a module in slot 1.

DIAGnostic:RELay:CYCLes:CLEar (@1201,1202)

Example: Resetting Module(s) to Power-On State

The following command

resets a module in slot 4 to its power- on state.

SYSTem:CPON 4

Installing SMA Connectors

When installing SMA connectors, it is recommend that you tighten them to
0.8 - 1.1 Nm (7- 10 in- lbs) of torque.