Trigger count, Trigger, Count – Agilent Technologies Switch/Measure User Manual

Page 61

background image

Features and Functions

2

34980A User’s Guide

49

Front Panel Operation:

Scan (Configure) > INTERVAL > EXTERNAL

To initiate the scan and store all readings in memory, press the
Scan (Measure) key. Between scan sweeps, “WAITING FOR TRIG” will be
displayed on the front panel. When a TTL pulse is received, the scan
starts and readings are stored in memory.

Note: To stop a scan, press and hold the

Scan (Measure) key.

Remote Interface Operation:

The following program segment configures the

instrument for an external scan.

Note: To stop a scan, press and hold the

Scan (Measure) key.

Trigger Count

You can specify the number of triggers that will be accepted by the
internal DMM before returning to the “idle” state. The trigger count applies
to both scanning and stand- alone DMM measurements (with no scan list).

Select a trigger count between 1 and 500,000 triggers, or continuous.

You can store at least 500,000 readings in memory and all readings are

automatically time stamped. If memory overflows, the new readings will
overwrite the first (oldest) readings stored; the most recent readings are
always preserved.

You can specify a trigger count in conjunction with a sample count and

a sweep count. The three parameters operate independent of one
another, and the total number of readings returned will be the product
of the three parameters.

The

CONFigure

and

MEASure?

commands automatically set the scan

trigger count to 1.

The instrument sets the scan trigger count to 1 after a Factory Reset

(

*RST

command). An Instrument Preset (

SYSTem:PRESet

command) or

Card Reset (

SYSTem:CPON

command) does not change the setting.

Front Panel Operation:

Scan (Configure) > SCAN TRIGGER > COUNTED|INFINITE

Remote Interface Operation:

TRIGger:COUNt

To configure a continuous scan, send

TRIG:COUNT INFINITY

.

TRIG:SOURCE EXT

TRIG:COUNT 2

INIT

Select external mode
Sweep the scan list 2 times
Initiate the scan