Multifunction Module with DIO, D/A, and Totalizer

10

34980A User’s Guide

199

34952A SCPI Programming Examples

The programming examples below provide you with SCPI command
examples to use for actions specific to the general purpose switch modules.

The slot and channel addressing scheme used in these examples follow the
form sccc where s is the mainframe slot number (1 through 8) and ccc is the
channel number. For information on specific configurations, refer to the
simplified schematic on

page 201

.

For complete information on the SCPI commands used to program the
34980A, refer to the Agilent 34980A Programmer’s Reference contained on
the 34980A Product Reference CD. For example programs, also refer to the
34980A Product Reference CD.

Digital Input/Output

Example: Configuring a DIO channel

The following program segment

configures channel 1 on the DAC module in slot 3 as an output and then
reads the output value (the channel is not reconfigured as an input). Then,
the channel is reconfigured as an input and the value is read again.

The second command below returns 64 as it is physically

reading the output data.

SOURce:DIGital:DATA:BYTE 64,(@3001)

SENSe:DIGital:DATA:BIT? 0,(@3001)

The second command below returns whatever is being input

externally.

CONFigure:DIGital:STATe INPut,(@3001)

SENSe:DIGital:DATA:BIT? 0,(@3001)

Totalizer

Example: Reading totalizer channel count

The following command reads the

count on totalizer channel 5 on the Multifunction module in slot 3.

SENSe:TOTalize:DATA? (@3005)

Example: Configuring the totalizer reset mode

To configure the totalizer reset

mode, send either of the following commands.

The following command configures totalizer channel 5 on the Multifunction
module in slot 3 to be read without resetting its count.

SENSe:TOTalize:TYPE READ,(@3005)