Autozero

When autozero is enabled (default), the instrument internally
disconnects the input signal following each measurement, and takes a
zero reading. It then subtracts the zero reading from the preceding
reading. This prevents offset voltages present on the instrument’s input
circuitry from affecting measurement accuracy.

When autozero is disabled, the instrument takes one zero reading and
subtracts it from all subsequent measurements. It takes a new zero reading
each time you change the function, range, or integration time.

Applies to temperature, dc voltage, resistance, temperature, and

dc current measurements only.

The autozero mode is set indirectly when you set the resolution and

integration time. Autozero is automatically turned off when you select
an integration time less than 1 PLC.

The

CONFigure

and

MEASure?

commands automatically enable autozero.

The autozero setting is stored in non- volatile memory, and does not

change when power has been off, after a Factory Reset (

*RST

command), or after an Instrument Preset (

SYSTem:PRESet

command).

Front Panel Operation:

DMM or Channel (Configure) > AUTO ZERO

Remote Interface Operation:

The OFF and ONCE parameters have a similar

effect. Autozero OFF does not issue a new zero measurement. Autozero
ONCE issues an immediate zero measurement.

[SENSe:]<function>:ZERO:AUTO {OFF|ONCE|ON} [,(@<ch_list>)]