Features and Functions

2

34980A User’s Guide

41

The

CONFigure

and

MEASure?

commands automatically set the gain

(“M”) to 1 and offset (“B”) to 0.

A Factory Reset (

*RST

command) turns off scaling and clears the

scaling values on all channels. An Instrument Preset (

SYSTem:PRESet

command) does not clear the scaling values and does not turn off scaling.

Front Panel Operation:

DMM or Channel (Configure) > SCALING > GAIN|OFFSET|UNITS

To define the label on the selected channel, press the arrow keys to move
the cursor to a specific position and then turn the knob to select the
desired letter or number. To clear the label on the selected channel,
change each character to “ ^ ” (starting with the rightmost character) and
then press the left arrow key to move to the next character.

Remote Interface Operation:

Use the following commands to set the gain,

offset, and custom measurement label.

CALC:SCALE:GAIN 1.2,(@1003)

CALC:SCALE:OFFSET 10,(@1003)

CALC:SCALE:UNIT 'PSI',(@1003)

After setting the gain and offset values, send the following command to
enable the scaling function on the specified channel.

CALC:SCALE:STATE ON,(@1003)