Features and Functions
2
34980A User’s Guide
23
Automatic Trigger Delays
If you do not specify a trigger delay, the instrument selects a delay
for you. The delay is determined by the function, range, integration time,
and ac filter setting as shown below.
DC Voltage, Thermocouple, DC Current (for all ranges):
Resistance, RTD, Thermistor (2- and 4-wire):
AC Voltage, AC Current (for all ranges):
Frequency, Period:
Digital Input, Totalize:
Integration Time
Trigger Delay
PLC > 1
PLC
≤ 1
2.0 ms
1.0 ms
Range
Trigger Delay
(for PLC > 1)
Range
Trigger Delay
(for PLC
≤ 1)
100
Ω
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
10 M
Ω
100 M
Ω
2.0 ms
2.0 ms
2.0 ms
25 ms
30 ms
200 ms
200 ms
100
Ω
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
10 M
Ω
100 M
Ω
1.0 ms
1.0 ms
1.0 ms
20 ms
25 ms
200 ms
200 ms
AC Filter
Trigger Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
7.0 seconds
1.0 second
120 ms
AC Filter
Trigger Delay
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
600 ms
300 ms
100 ms
Trigger Delay
0 seconds