beautypg.com

Automatic trigger delays – Agilent Technologies Switch/Measure User Manual

Page 35

background image

Features and Functions

2

34980A User’s Guide

23

Automatic Trigger Delays

If you do not specify a trigger delay, the instrument selects a delay
for you. The delay is determined by the function, range, integration time,
and ac filter setting as shown below.

DC Voltage, Thermocouple, DC Current (for all ranges):

Resistance, RTD, Thermistor (2- and 4-wire):

AC Voltage, AC Current (for all ranges):

Frequency, Period:

Digital Input, Totalize:

Integration Time

Trigger Delay

PLC > 1
PLC

≤ 1

2.0 ms
1.0 ms

Range

Trigger Delay
(for PLC > 1)

Range

Trigger Delay
(for PLC

1)

100

1 k

10 k

100 k

1 M

10 M

100 M

2.0 ms

2.0 ms
2.0 ms
25 ms
30 ms
200 ms
200 ms

100

1 k

10 k

100 k

1 M

10 M

100 M

1.0 ms

1.0 ms
1.0 ms
20 ms
25 ms
200 ms
200 ms

AC Filter

Trigger Delay

Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)

7.0 seconds
1.0 second
120 ms

AC Filter

Trigger Delay

Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)

600 ms
300 ms
100 ms

Trigger Delay

0 seconds