beautypg.com

Agilent Technologies Switch/Measure User Manual

Page 141

background image

Low Frequency Multiplexer Switch Modules

4

34980A User’s Guide

129

using the *RST command. This command resets the mainframe and all

installed modules to the Factory configuration. This affects all
installed modules.

cycling system power. This affects all installed modules.

If the overvoltage situation is not resolved, clearing the overvoltage will
result in a new overvoltage event occurring immediately.

Further FET protection is assured only as one channel in each bank is
closed at any time. Thus this module will operate as only a 1:N MUX
module. For more information about FET channel closures, refer to

page 98

.