Agilent Technologies Switch/Measure User Manual
Page 23
Features and Functions
2
34980A User’s Guide
11
The Analog Bus relays (numbered s911, s912, s913, etc.) on the
multiplexer and matrix modules are ignored if they are included in a
range of channels. An error will be generated if an Analog Bus relay is
specified as the first or last channel in a range of channels. For example,
the following command closes all valid channels between channel 30
(slot 1) and channel 5 (slot 2). In addition, this command closes Analog
Bus relay 911 on the module in slot 1 (Bank 1). Note that although the
specified range of channels includes the other Analog Bus relays, they are
ignored and are not closed by this command.
ROUT:CLOS (@1030:2005,1911)
The following command will generate an error since the Analog Bus relays
cannot be specified as the first or last channel in a range of channels
(none of the channels will be closed).
ROUT:CLOS (@1005:1911) !Generates an error
In the following command, since the optional
omitted, the command will be applied to the internal DMM. If the internal
DMM is disabled or is not present, an error will be generated.
INP:IMP:AUTO ON !Applies to the internal DMM