Configurable Parameters
Table 47 below lists the configurable parameters and their
allowed and default values for the WS425 A/B NMEA Extended
profile.
Table 47
Configurable Parameters for WS425 A/B
NMEA Extended Profile
Parameter
Default
Value
Profile-Specific
Allowed Values
Description
address
A
String with 1 character
Address for WMT700
autoInt
1
0.25 ... 1000
Resolution: 0.25
Automatic message interval in seconds. Do
not select a message interval that is shorter
than the time it takes to send a data message
autoPort
2
1 = COM1 port
2 = COM2 port
Serial port to which WMT700 sends automatic
data messages
autoSend
15
0 = Automatic
messages disabled
15 = NMEA automatic
data message
Automatic data message number. Selects the
data message format for automatic messages
com1_protocol 0*
5 = WS425 A/B NMEA
Extended
Profile for serial port COM1
com2_protocol
5
5 = WS425 A/B NMEA
Extended
Profile for serial port COM2
com2_interf
N/A**
0 = RS-485
1 = RS-422
3 = RS-232
Interface for serial port COM2
wndAvg
3
0.25 … 3600
Resolution: 0.25
Averaging time for wind measurement in
seconds
wndOrientation 0
0 = Array facing up
1 = Array facing down
Orientation of the array of WMT700
wndUnit
0**
0 = Meters per second
1 = Miles per hour
2 = Kilometers per hour
3 = Knots
Wind speed unit
wndVector
0
0 = Scalar averaging
Wind averaging method
* COM1 service port is always by default 0 - WMT700 protocol. This can be
changed using serial commands if COM1 is needed for protocol-specific
communication.
** Wind speed unit and COM2 digital communication interface are defined in
the order form. Default setting may be any of the allowed options, and it can be
verified from the unit’s configuration code.
156 __________________________________________________________________ M211095EN-E