Chapter 7 ____________________________________________________________ Technical Data

VAISALA_______________________________________________________________________ 75

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the GMP343.

Performance

Sensor

Vaisala CARBOCAP

®

Measuring principle

Single-Beam Dual-Wavelength NDIR

Measurement range options

0 ... 1000 ppm
0 ... 2000 ppm
0 ... 3000 ppm
0 ... 4000 ppm
0 ... 5000 ppm
0 ... 2 %

Accuracy (excluding noise) at 25 °C and 1013 hPa after factory calibration with 0.5 % accurate
gases with different range options

0 ... 1000 ppm
0 ... 2000 ppm - 0 ... 2 %*

±(3 ppm CO

2

+ 1 % of reading)

±(5 ppm CO

2

+ 2 % of reading)

* Accuracy below 200 ppm CO

2

not specified for the 2 % range option

Noise (repeatability) at 370 ppm CO

2

with no output averaging
with 30 s output averaging

±3 ppm CO

2

±1 ppm CO

2

Short-term stability (up to 6 hours) at 370 ppm CO

2

±1 ppm CO

2

Long-term stability

In easy operating conditions
In moderate oper. conditions
In harsh operating conditions

±2 % of reading */ year
±2 % of reading */ 6 months
±2 % of reading */ 3 months

* Always at least ±10 ppm CO

2

See

Figure 12 on page 79

.