Preparation for Network Operation

Before GMP343 is connected to the network it must be configured. Normal
procedure is shown below:

First connect the device to PC by using RS-232 and when communication
is currently open, key in the following commands:

RSMODE 485

ADDR addr

(where addr is 1...99)

SMODE POLL

SAVE

Next turn the power off. Then you can connect the device to RS-485
network.

Commands during Network Operation

When GMP343 is connected to the network, the following commands can
be entered:

Compensation values that are entered with compensation commands (XP,
XO, XRH) are used for compensation but they are not stored permanently.
Note that compensation commands do not give any reply. If value in
command is not acceptable, the command is rejected.

SEND addr

Return latest CO

2

value

XP addr pressure

Set pressure for compensation

XO addr oxygen

Set oxygen for compensation

XRH addr
humidity

Set humidity for compensation

OPEN addr

Open device for normal communication (i.e. to
STOP-mode)

CLOSE

Switch the device from STOP-mode to POLL-
mode.