beautypg.com

Open and close, Setting echoing mode, Open and close setting echoing mode – Vaisala GMP343 User Manual

Page 40

background image

User’s Guide ______________________________________________________________________

38 ___________________________________________________________________M210514EN-E

Open and Close

OPEN nn

nn=address of the transmitter (0...99)

The OPEN command sets the bus temporarily in STOP mode so that the
SMODE command can be entered.

CLOSE

In STOP mode: Command OPEN has no effect, CLOSE sets the
transmitter temporarily in POLL mode.

In POLL mode: Command OPEN sets the transmitter temporarily in
STOP mode.

Example:

>close

line closed

GMP343: 1 line opened for operator commands

>send

351.1 ppm

>smode stop

SMODE : STOP

>

Setting Echoing Mode

ECHO x

SAVE

x=ON/OFF (default ON)

In RS-232 mode the device echoes everything back to the user. In RS-485
mode echoing is automatically disabled.