Error states, Table 7, Troubleshooting errors – Vaisala GMP343 User Manual
Page 74
User’s Guide ______________________________________________________________________
72 ___________________________________________________________________M210514EN-E
Error States
In case of an error function the outputs are as follows:
-
Serial line: In STOP mode the error code and description (see the
Table 6 below) is shown when keying in the command: ERRS. In
POLL or RUN mode the error flag is set to 1 (if the error flag is
included in the message format, see FORM-command). Reset the
error state with a command ERRS R.
-
Analog output shows the error level (can be set by using the AERR-
command. Factory setting: 5 V or 20 mA).
Table 7
Troubleshooting Errors
Error code
Error description
Interpretation
Action
E01
EEPROM checksum failure Internal transmitter failure
Return the device to
Vaisala Service
E02
IR source failure
Internal transmitter failure
Return the device to
Vaisala Service.
E03
FPI failure
Internal transmitter failure
Return the device to
Vaisala Service.
E04, E05
Heater failure
Internal transmitter failure
Return the device to
Vaisala Service.
E06
Temperature measurement
failure
Operation temperature is
out of allowed range.
Analog output: Error level
is shown if temperature
compensation is enabled.
Otherwise, output is
normal.
Ensure that the operating
temperature is -45...+85
°C (-49...185 °F). In case
of constant error, return
the device to Vaisala
Service.
E07
Measurement signal level
too low
Measurement chamber is
contamined or the lamp is
degraded.
Change the filter and clean
the optics according to the
instructions in section
Changing the Filter and
Cleaning the Optics
(Diffusion Model Only) on
page 70
. In case of
constant error, return the
device to Vaisala Service.
W01
Watchdog reset occured
Software defect
In case this warning
appears frequently, return
the device to Vaisala
Service.
W02
Stack overflow
Software defect
In case this warning
appears frequently, return
the device to Vaisala
Service.