beautypg.com

Setting the analog output overrange clip, Testing the analog output – Vaisala GMP343 User Manual

Page 57

background image

Chapter 4 ________________________________________________________________ Operation

VAISALA_______________________________________________________________________ 55

Setting the Analog Output Overrange Clip

This command clips the analog output signal at the high value when the
signal goes over range. In case of the current output the current will not go
higher than 20 mA or below the low value current (set with ILOW). The
voltage output will not go higher than the voltage set with UHIGH.

ACUT x

SAVE

x=ON/OFF

>acut

ACUT : OFF ? ON

>acut

ACUT : ON ? OFF

>

>save

EEPROM saved successfully.

>

Testing the Analog Output

When you want to test the operation of the analog output use this command
to force the output to the set value. The value in the analog output can then
be measured with a current/voltage meter. The set value remains valid until
you give the ATEST command without a value or RESET the transmitter.

ATEST

x=the given test value (mA or V)

>atest 1

Test voltage set at 1 V. Use command without any parameters

to stop test mode.

>atest

Voltage test mode stopped.