How the filtering affects the response time, Setting the median filter, Figure 9 – Vaisala GMP343 User Manual
Page 45: The difference of the two filters in time response
Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 43
How the Filtering Affects the Response Time
The figure on the next page illustrates the difference of the two filters in
regard to time response. The averaging time is set to 40 seconds while the
smoothing factor is set to 10 to have about equal noise rejection properties.
See also the response time tables presented in Specifications
.
Figure 9
The Difference of the Two Filters in Time Response
Setting the Median Filter
Use the median filter to remove random peak values caused by possible
external interference.
MEDIAN x
SAVE
x=0...13 (default=0)
>median 3
MEDIAN : 3
>save
EEPROM saved successfully (829 ms).
>