Setting the Analog Output Error Level

Factory default states for analog outputs during error condition are the
maximum values of the measuring range=2.5 V / 5 V / 20 mA.

AERR x

SAVE

x=error value (mA) or (V)

Acceptable ranges: 0...20 mA, 0...6 V

Example:

>aerr

AERR (V) : 5.000 ? 0

>

>save

EEPROM saved successfully.

>

NOTE

The error output value must be within a valid range for current (0...20
mA) or voltage output (0...6 V) regardless of the UHIGH setting.