Setting the Continuous Outputting
Interval
INTV xxxx yyy
SAVE
xxxx=output interval (1...1000) default=1 s
yyy=unit (s, min or h)
Example: Output interval is changed to 5 seconds
>intv 5 s
INTERVAL : 5
UNIT : S
>
Outputting the Reading Once
In STOP mode:
SEND
In POLL mode:
SEND aa
aa=address of the transmitter when more than one transmitter is connected
to a serial bus (0...99).
The output mode can be changed with the command FORM.
>send
348.7 ppm
>
Setting the Serial Interface Measurement
Mode
SMODE xxxx
SAVE
xxxx= STOP/RUN/POLL
See also other documents in the category Vaisala Tools:
- DM500 (138 pages)
- DM70 (93 pages)
- DMT132 (74 pages)
- DMT143 (76 pages)
- DMT152 (70 pages)
- DMT242 (4 pages)
- DMT340 (191 pages)
- DMT345 (185 pages)
- DPT145 (63 pages)
- DPT146 (71 pages)
- PTU300 (217 pages)
- PTB330TS (89 pages)
- PTB220 (10 pages)
- PTB220 (113 pages)
- PTB330 (144 pages)
- PTU200 (64 pages)
- PTU200MIK1 (18 pages)
- SPH10 (2 pages)
- SPH20 (2 pages)
- PTB110 (4 pages)
- PTB200 (30 pages)
- PTB210 (analog) (27 pages)
- PTB210 (serial) (32 pages)
- GM70 (68 pages)
- GMD20 (4 pages)
- GMK220 (18 pages)
- GML20 (2 pages)
- GML20T (2 pages)
- GMM20W (5 pages)
- GMM220 (6 pages)
- GMP231 (2 pages)
- GMP231 (90 pages)
- GMR20 (2 pages)
- GMT220 (42 pages)
- GMW90 (101 pages)
- XMW90 (4 pages)
- MM70 (67 pages)
- MM70 (71 pages)
- MMT162 (66 pages)
- MMT310 (81 pages)
- MMT330 (171 pages)
- MMT330 (181 pages)