beautypg.com

Setting the averaging filter, Setting the smoothing filter, Flushing the filters (resynchronization) – Vaisala GMP343 User Manual

Page 46

background image

User’s Guide ______________________________________________________________________

44 ___________________________________________________________________M210514EN-E

Setting the Averaging Filter

Use the averaging filter to reduce the noise of the measurement. Notice the
effect on the response time.

AVERAGE x

SAVE

x=0...60 seconds (default =10 s)

>average 20

AVERAGING (s) : 20

>save

EEPROM saved successfully (829 ms).

>

Setting the Smoothing Filter

Use the smoothing filter to reduce the noise of the measurement for long
averaging periods. Notice the effect on the response time.

SMOOTH x

SAVE

x=0...255 (default =0)

>smooth 10

SMOOTH : 10

>save

EEPROM saved successfully (829 ms).

>

Flushing the Filters (Resynchronization)

With this command you can flush the filters to get rid of the effect of the
past measurements. This feature is useful if long filtering lengths are used.

RESYNC

>resync

>