Setting the High Limit of the
Concentration Range

Sets the hight CO

2

concentration value, which corresponds to the analog

output high value.

AHIGH

SAVE

Key in the highest concentration value (ppm) and press ENTER.

>ahigh

AHIGH (ppm) : 1000.0 ? 1200

>

>save

EEPROM saved successfully.

>

Setting the Low Limit of the
Concentration Range

Sets the low CO

2

concentration value, which corresponds to the analog

output low value.

ALOW

SAVE

Key in the lowest concentration value (ppm) and press ENTER.

>alow

ALOW (ppm) : 0.0 ? 20

>

>save

EEPROM saved successfully.

>