Chapter 4 ________________________________________________________________ Operation

VAISALA_______________________________________________________________________ 31

Getting the Measurement Message

Measurement Units

GMP343 outputs the following units:

-

Carbon dioxide (ppm)

-

Temperature (ºC)

For other measurement units (%, non-metric units), use the Vaisala MI70
indicator.

Starting the Continuous Outputting

R

This command starts the continuous outputting of results (output form
defined with the command FORM). Data is printed with the interval
defined with the command INTV. Printing can be stopped with the
command S.

>r 345.0 ppm

344.1 ppm

343.6 ppm

345.6 ppm

346.1 ppm

344.1 ppm

343.5 ppm

345.5 ppm

>

Stopping the Continuous Outputting

S

Ends the RUN mode; after this command all other commands can be used.

NOTE

The built-in temperature sensor is intended for compensation of the CO

2

measurement. By default the compensation is enabled. For more
information, see section

Temperature, Pressure, Relative Humidity, and

Oxygen Compensations on page 45

.

NOTE

Measurement units (ppm and %) express the CO

2

concentration by gas

volume.1 % CO

2

= 10 000 ppm CO

2