beautypg.com

1installation, R3 r6 – Vaisala HMP260 User Manual

Page 91

background image

HMP260 SERIES
Appendix 3: RS 485/422 serial port module

HMP260-U017en-1.5

1

INSTALLATION

Switch the transmitter off.

Resistors R2, R3 and R4 between connectors X4 and X5 in the component
board in the cover of the transmitter are removed with side-cutting pliers. The
module is plugged in connectors X4 and X5 on the main board of the
HMP260 transmitter; connector X1 on the module board to connector X4 and
connector X2 to connector X5.

X1

X2

24 V

+ -

C h1

Ch2

-

- +

+

RX

GND TX

R 2

R3

R4

X6

RX H I

RX

LO

TX

LO

TX

HI

X4

X5

RS4 8 5/ 4 22

RX

TX

X1

X2

H

I

H

I

LO

L

O

X1

X2

R3

R6

Jumper ( ) selections for the
RS 485/422 serial bus module

Single pair

Dual pair

serial bus

module

Cut off the resistors R2, R3 and R4 on the main board.
New signal names for X6 screw terminal are on the module.
Follow the instructions on the module:
X1 to X4 and
X2 to X5 on the mother board

NOTE! If the transmitter is NOT at the end
of the bus OR the line has a dynamic line
termination, the resistors R3 and R6 have
to be removed!

Connect the data wires to screw terminal X6 on the main board. Switch the power on.