beautypg.com

8 electromagnetic compatibility, 1 emissions, 2 immunity – Vaisala HMP260 User Manual

Page 64

background image

HMP260 SERIES
Operating Manual

HMP260-U017en-1.5

58

1997-11-25

8.8

Electromagnetic compatibility

The emission and immunity tests have been performed according to standards
EN50081-1 and EN50082-2.

NOTE

These tests are valid with specified sensor protections
only.

Type:

Sensor protection/notes:

HMP263

PPS grid & stainless steel netting
(part no. 16720)

HMP264

sintered filter (part no. 16452)

HMP265

sintered filter (part no. 16452), PPS grid &
stainless steel netting (part no. 16720)

8.8.1

Emissions

Test:

Setup according to:

Radiated interference

EN55022 (class B)

8.8.2

Immunity

Test:

Setup according to:

Performance:

Electrostatic discharge

EN61000-4-2:1995

criteria B

Electrical fast transients

EN61000-4-4:1995

criteria B

GSM field immunity

ENV50204:1995

criteria A

RF radiated fields

ENV50140:1993

criteria A

RF conducted fields

ENV50141:1993

criteria A

Voltage proof, AC: DC supply (+ or -) to housing 250 VAC, 1 minute (300 k

and 15 nF parallel)