8 electromagnetic compatibility, 1 emissions, 2 immunity – Vaisala HMP260 User Manual
Page 64

HMP260 SERIES
Operating Manual
HMP260-U017en-1.5
58
1997-11-25
8.8
Electromagnetic compatibility
The emission and immunity tests have been performed according to standards
EN50081-1 and EN50082-2.
NOTE
These tests are valid with specified sensor protections
only.
Type:
Sensor protection/notes:
HMP263
PPS grid & stainless steel netting
(part no. 16720)
HMP264
sintered filter (part no. 16452)
HMP265
sintered filter (part no. 16452), PPS grid &
stainless steel netting (part no. 16720)
8.8.1
Emissions
Test:
Setup according to:
Radiated interference
EN55022 (class B)
8.8.2
Immunity
Test:
Setup according to:
Performance:
Electrostatic discharge
EN61000-4-2:1995
criteria B
Electrical fast transients
EN61000-4-4:1995
criteria B
GSM field immunity
ENV50204:1995
criteria A
RF radiated fields
ENV50140:1993
criteria A
RF conducted fields
ENV50141:1993
criteria A
Voltage proof, AC: DC supply (+ or -) to housing 250 VAC, 1 minute (300 k
Ω
and 15 nF parallel)