4checking the serial port network operation – Vaisala HMP260 User Manual
Page 108

HMP260 SERIES
Appendix 4: Digital current loop module
HMP260-U017en-1.5
4
CHECKING THE SERIAL PORT NETWORK OPERATION
Normally measurement readings are asked when the transmitter is in POLL mode; the
command must then include the address of the transmitter. If the settings of the
transmitter need to be changed, the transmitter is switched to STOP mode with
command OPEN; the commands can then be given without address. When the line to
the transmitter is closed, it returns to POLL mode.
STOP mode
Open the line to the transmitter:
open 22
HMP 22 line opened for operator commands
Transmitter no. 22 is now temporarily set to STOP mode; it accepts commands without
address until CLOSE command is given. Individual settings can now be easily modified.
Do not open more than one line at a time.
Use command ? to find out the settings of the active transmitter:
?
HMPS260 / 1.01
CPU serial nr
: 0
Keyboard type
: 0
Address
: 7
Output units
: metric
Baud P D S
: 4800 E 7 1 FDX
Serial mode
: STOP
Output intrv.
: 0 min
Mtim
: 32
Pressure
: 1013.25
Analog outputs
Ch1 0.00 ... 10.00 V
Ch2 0.00 ... 10.00 V
Ch1 ( RH ) lo
0.000 %RH
Ch1 ( RH ) hi 100.000 %RH
Ch2 ( T ) lo -20.000 'C
Ch2 ( T ) hi 180.000 'C
Transducer
:
PRB serial nr
: 0
Calibr. date
: 0
When the necessary settings have been made, close the line to transmitters (the
command closes all open lines):
>close
line closed
CLOSE command is always given without address. If there was no line open, there will
be no response to the CLOSE commands.