Appendix 1: Serial commands
The output scales can also be given directly with the ASCL command.
ASCL aaa.a bbb.b ccc.c ddd.d
aaa.a
=
lower limit of channel 1
bbb.b
=
upper limit of channel 1
ccc.c
=
lower limit of channel 2
ddd.d
=
upper limit of channel 2
For example, when relative humidity is scaled to 0...100 %RH on channel 1 and
temperature to -40...+100
°
C on channel 2:
ASCL 0 100 -40 100
Ch1
(RH)
lo
0.000 %RH
Ch1
(RH)
hi
100.000 %RH
Ch2
(T ) lo
-40.000 'C
Ch2
(T ) hi
100.000 'C
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP240 (130 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)