6.3

Calibration of analogue outputs

The analogue outputs have been calibrated at the factory and since they are
very stable, calibration of the outputs should be performed only when there is
reason to believe that their adjustments have changed.

6.3.1

Using serial commands

Disconnect the security lock jumper!

ACAL

The outputs on channels 1 and 2 are measured and the measured values (mA
or V) entered as calibration coefficients.

Example: both channels have 0...10 V outputs (set with AMODE command);
enter the voltages measured at the analogue outputs:

>ACAL
Ch1

U1 ( V ) ? 0.123

Ch1

U2 ( V ) ? 9.98

Ch2

U1 ( V ) ? 0.120

Ch2

U2 ( V ) ? 9.98

6.3.2

Using display/keypad commands

Disconnect the security lock jumper!

Connect an ammeter/voltmeter to the output of channel 1, select

Cali

in the main menu and

Analog outputs in the Cali menu. The follow-

ing is displayed (the quantity can be either mA or V):

Enter the measured lower end current/voltage on channel 1.

Enter the measured upper end current/voltage on channel 1.