Appendix 4: Digital current loop module

POLL mode

If a transmitter has been set to POLL mode, it will respond only to commands sent with
its address:

send 22
RH= 24.4 %RH T= 29.1 'C

Addresses from 1 to 99 can be used. According to the 20 mA current loop standard
current flows with no transmission on line. A maximum of 6 devices can be connected
on same single loop line, but the number can be raised to 9 by using dual loop wiring.

Current loop settings

HMP260 settings

single pair

dual pair

Full duplex/half duplex

HDX

FDX

Echo on/off

OFF

ON

Terminal settings
Line feed after carriage return

yes

no

HDX/FDX

FDX

FDX

When terminal is set to general (line feed) after (carriage return), the listings
will have two line feeds where also the HMP260 transmitters send line feed.