HMP260 SERIES
HMP260-U017en-1.5
Operating Manual
1997-11-25
33
5.4.1.2
Selecting and scaling the analogue output quantities
Disconnect the security lock jumper!
ASEL xxx yyy
xxx
=
channel 1's quantity
yyy
=
channel 2's quantity (RH, T, Td, Abs, Mix or Tw)
Example:
relative humidity selected on channel 1 and temperature on channel 2
ASEL RH T
Ch1 (RH)
lo
0.000 %RH ?
Ch1 (RH)
hi
100.000 %RH ?
Ch2 (T )
lo
-40.000 'C
?
Ch2 (T )
hi +160.000 'C
?
5.4.1.3
Scaling the analogue outputs
Disconnect the security lock jumper!
ASCL
Example:
relative humidity is scaled on the range of 0...100 %RH and temperature -
40...+160 °C
ASCL
Ch1 (RH)
lo
0.000 %RH ?
Ch1 (RH)
hi 100.000 %RH ?
Ch2 (T )
lo
0.000 'C
? -40
Ch2 (T )
hi 100.000 'C
? 160
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP240 (130 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)