One point calibration procedure, 1 using serial commands, 2 using display/keypad commands – Vaisala HMP260 User Manual
Page 45
![background image](/manuals/408447/45/background.png)
HMP260 SERIES
HMP260-U017en-1.5
Operating Manual
1997-11-25
39
NOTE
If the sensor has been changed, the calibration has to be
done according to the instructions in Chapter 6.1.3.
6.1.1.
One point calibration procedure
6.1.1.1
Using serial commands
Disconnect the security lock jumper!
•
Make sure that the sensors of the transmitter and the reference in-
strument are close to each other. Allow enough time for the sensor
heads to stabilize to the measurement conditions
•
Give command CRH
>CRH
RH :
xx.x
Ref1 ?
yy.y
Press any key when ready...
•
If you want to see how the sensor stabilizes to the reference humidity
enter c
RH : 11.9
Ref1
? c
RH : 11.5
Ref1
? c
RH : 11.5
Ref1
? 11.3
Press any key when ready...
•
Press any key and press
point value.
RH
:
yy.y
Ref2 ?
6.1.1.2
Using display/keypad commands
Disconnect the security lock jumper!
•
Make sure that the sensors of the transmitter and the reference in-
strument are close to each other. Allow enough time for the sensor
heads to stabilize to the measurement conditions
•
Select
Cali in the main menu and then RH cal; select Not changed
and then one-point offset correction
RH 1 point cal. Change the
humidity reading with the arrow keys to correspond to the reference
value and acknowledge it with ENT; pressing an arrow once changes
the reading by 0.05 %RH.