Page 52

background image

HMP260 SERIES
Operating Manual

HMP260-U017en-1.5

46

1997-11-25

6.2.1.3

Using LED commands

Disconnect the security lock jumper!

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

Check the transmitter against the reference.

Connect an ammeter/voltmeter to the analogue outputs (connector
X2). Give command

¡¡l¡. At the first calibration point the LED

on the left flashes; adjust the first point (offset) with the arrow
switches to the same reading with the reference and press ENT
switch.

After adjusting the offset point and pressing ENT the second LED
from left flashes. Press ENT without changing the output value.

6.2.2

Two point temperature calibration

6.2.2.1

Using serial commands

Disconnect the security lock jumper!

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

Check the transmitter against the reference.

Give command CT and enter the first point value and press
:

CT
'T : xx.x Ref1 ? yy.y
Press any key when ready

If you want to see how the sensor stabilizes to the reference
temperature, enter c instead of the first reference:

T : 0.90 Ref1 ? c
T : 0.55 Ref1 ? c
T : 0.55 Ref1 ? 0.0
Press any key when ready...

Change the temperature and check the transmitter again against the
reference.

Check that the reading corresponds to the reading of the reference
instrument. If not, adjust the second point

Press any key, enter the second point value and press .

'T : xx.x Ref2 ? yy.y