HMP260 SERIES

HMP260-U017en-1.5

Operating Manual

1997-11-25

41

If you want to see how the sensor stabilizes to the humidity in the
calibrator, enter c instead of the first reference:

RH : 11.9 Ref1 ? c
RH : 11.5 Ref1 ? c
RH : 11.5 Ref1 ? 11.3
Press any key when ready...

Place the sensor head into the calibration hole of the NaCl bottle in
the humidity calibrator.

Wait for 10 minutes.

Press any key and enter the second point value and press .

RH : xx.x Ref2 ? yy.y

The stabilization of the sensor can be monitored by entering c
instead of the reference value.

6.1.2.2

Using display/keypad commands

Disconnect the security lock jumper!

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

Place the sensor head into the calibration hole of the LiCl bottle in the
humidity calibrator.

Wait for 10 minutes.

Select Cali in the main menu and then RH cal; select Not changed
and then two-point calibration RH 2 point cal. Change the first point
reading with the arrow keys and press ENT.

Place the sensor head into the calibration hole of the NaCl bottle in
the humidity calibrator.

Wait for 10 minutes.

If necessary, change the second point reading with the arrow keys and
press ENT.

6.1.2.3

Using LED commands

Disconnect the security lock jumper!

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

Place the sensor head into the calibration hole of the LiCl bottle in the
humidity calibrator.