4 setting the output interval for the run mode, 5 serial bus settings – Vaisala HMP260 User Manual
Page 41

HMP260 SERIES
HMP260-U017en-1.5
Operating Manual
1997-11-25
35
5.4.2.4
Setting the output interval for the RUN mode
INTV xxx yyy
xxx
=
output interval (0...255)
0: no pause between outputs
yyy
=
unit (s, min or h)
Example:
output interval is changed into 10 minutes
INTV 10 min
Output intrv. : 10 min
5.4.2.5
Serial bus settings
SERI b p d s x
b
=
bauds (300, 600, 1200, 2400, 4800, 9600)
p
=
parity (n = none, e = even, o = odd)
d
=
data bits (7 or 8)
s
=
stop bits (1 or 2)
x
=
duplex (H = half, F = full)
The processor does not allow the following combinations:
•
no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP260 programme will change the number of stop bits to 2
•
even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
The settings can be changed one parameter at a time or all parameters at once:
SERI O
changing parity only
4800 O 7 1 HDX
SERI 600 N 8 1 F
changing all parameters
600 N 8 1 FDX
When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.