beautypg.com

Vaisala HMP260 User Manual

Page 73

background image

HMP260 SERIES
Appendix 1: Serial commands

HMP260-U017en-1.5

In one-point offset correction, the Ref2 prompt is acknowledged with :

CT

T : 0.80 Ref1 ?

0.0

Press any key when ready ...

T : 75.50

Ref2 ?

If the stabilization of the sensor to the temperature of the calibrator or the reference
needs to be monitored, the measurement output can be repeated by giving command
c at Ref1 and Ref2:

CT

T : 0.80 Ref1 ?

c

T : 0.40 Ref1 ?

0.00

Press any key when ready ...

T : 56.20

Ref2 ?

55.0

ACAL

Calibrating the analogue outputs

Disconnect the security lock jumper!

ACAL

Calibrates the outputs selected on channels 1 and 2. The output is measured and the
measured values (mA or V) entered as calibration coefficients.

For example, calibrating the outputs when 0...10 V signal has been selected on both
channels (set with AMODE command)

ACAL

Ch1

U1

(V ) ?

0.123

Ch1

U2

(V ) ?

9.98

Ch2

U1

(V ) ?

0.120

Ch2

U2

(V ) ?

9.98

L

Outputting linear correction coefficients

L

With the help of command L the user can check how the transmitter has been
adjusted after it has been calibrated at the factory.

L

RH

offset

: 0.000

RH

gain

: 1.000

T

offset

: 0.000

T

gain

: 1.000