HMP260 SERIES
Appendix 1: Serial commands

HMP260-U017en-1.5

2

CALIBRATION COMMANDS

CRH

Relative humidity calibration

Disconnect the security lock jumper!

CRH

With this command the transmitters can be calibrated against a reference. A one-point
calibration can be done against an accurate transfer standard in the field and a two-point
calibration using saturated salt solutions in controlled conditions. A two-point calibration
is performed as follows:

CRH

RH : 12.00 Ref1 ? 11.3

Press any key when ready ...

RH : 76.00 Ref2 ? 75.5

In one-point offset correction, the Ref2 prompt is acknowledged with :

CRH

RH : 12.80 Ref1 ? 11.3

Press any key when ready ...

RH : 75.50 Ref2 ?

If the stabilization of the sensor to the humidity in the calibrator needs to be monitored,
the measurement output can be repeated by giving command c at Ref1 and Ref2:

CRH

RH : 12.00

Ref1 ? c

RH : 11.70

Ref1 ? c

RH : 11.50

Ref1 ? 11.3

Press any key when ready ...

RH : 76.00

Ref2 ? 75.5