2calibration commands – Vaisala HMP260 User Manual
Page 71

HMP260 SERIES
Appendix 1: Serial commands
HMP260-U017en-1.5
2
CALIBRATION COMMANDS
CRH
Relative humidity calibration
Disconnect the security lock jumper!
CRH
With this command the transmitters can be calibrated against a reference. A one-point
calibration can be done against an accurate transfer standard in the field and a two-point
calibration using saturated salt solutions in controlled conditions. A two-point calibration
is performed as follows:
CRH
RH : 12.00 Ref1 ? 11.3
Press any key when ready ...
RH : 76.00 Ref2 ? 75.5
In one-point offset correction, the Ref2 prompt is acknowledged with
CRH
RH : 12.80 Ref1 ? 11.3
Press any key when ready ...
RH : 75.50 Ref2 ?
If the stabilization of the sensor to the humidity in the calibrator needs to be monitored,
the measurement output can be repeated by giving command c
CRH
RH : 12.00
Ref1 ? c
RH : 11.70
Ref1 ? c
RH : 11.50
Ref1 ? 11.3
Press any key when ready ...
RH : 76.00
Ref2 ? 75.5