beautypg.com

Main board hmp230 – Vaisala HMP260 User Manual

Page 116

background image

BLU

GRN

YEL

BLK

HMP 265

In automation system
galvanically isolated inputs
are recommended for current signals

Main board HMP230

To control system

Supply voltage 24 VDC
max. 120 mA

Note ! The cable shield has to be connected to the cable bushing for full EMC protection.

Process temperature

Relative humidity

(To control system)

X2

1

2

3

4

1

2

3

1

2

4

Channel1+

Channel1-

Channel2+

Channel2-

RX

GND

TX

NC

-

+

Serial
connections
See manual for
details!

Humidity and
temperature outputs
current/voltage
Max. load for current
outputs 500 ohm

Probe cable length 2, 5 or 10 m

X6

X1

Process
temperature
4-wire connection
Pt 100

RH sensor
HUMICAP

GND

RH2

RH1

Drawn

Review

Appr

Design

Replaces

Scale

id

Title

Arch

no

Serial

Sheet

doc no

Cooperator`s

Dwg no

Rev

Ltr

Qty

Change

Reason/

ECO no

Design

Date Review

Date Appr

Tool

Qty

pcs

+RX

-RX

+RX

-RX

TX

RX

RS 232

Digital cur-
rent loop

RS 485

HI

LO

HI

LO

94-02-16 KKe

KKe

HMP263, HMP264 AND HMP265
WIRING DIAGRAM

Hazardous area
EEx ia IIC T3/T4

Protection unit HMP260PU
[EEx ia] IIC

Housing of the HMP260

Location of main board terminals

X2

1 2 3 4

4
3
2
1

2 1

X1

X6

X15

Jumper selections in connector X15

Current 0...20 mA/4...20 mA

Ch1

C h 2 C h 1

Ch2

Ch1

C h 2 C h 1

Ch2

Ch1

C h 2 C h 1

Ch2

Voltage 0...1 V

See manual for details!

Voltage 0...5 V/0...10 V

BLU

GRN

YEL

BLK

HMP 263

BLU

GRN

YEL

BLK

HMP 264