beautypg.com

Vaisala HMP260 User Manual

Page 82

background image

HMP260 SERIES

HMP260-U017en-1.5

Appendix 1: Serial commands

ADDR

Setting the transmitter address

ADDR aa

aa

=

address (0...99)

The address is used when more than one transmitter is connected to one serial bus. The
ADDR command makes it possible to communicate with one transmitter at a time in
POLL mode.

For example, transmitter is given address 99

ADDR

Address : 2 ? 99

When asking the current address, no address number is given:

ADDR

Address : 2 ?

RESET

Resetting the transmitter

RESET

Resets the transmitter. All settings that have been changed stay in the memory even after
reset or power failure.

3.1

Operation modes

SMODE

Setting the serial interface

SMODE xxxx

xxxx

=

STOP, RUN or POLL

In STOP mode:

measurements output only by command, all commands can be used

In RUN mode:

outputting automatically, only command S can be used

In POLL mode:

measurements output only with command SEND. When in POLL
mode, the output mode is changed as follows: